Szukaj

    Nakanishi Inc.NSK Logo

    Rozwiązywanie problemów

    Higiena jamy ustnej

    Varios Combi Pro2

    • A

      Error indication

      Cause of error

      Wireless communication disconnected during ultrasonic mode or powder mode.

      Release method

      Touch

      Error indication

      Cause of error

      Wireless communication disconnected during ultrasonic mode or powder mode.

      Release method

      Depress the ON/OFF button.

      Error indication

      "Battery" is displayed and is flashing.

      Cause of error

      "BATTERY" is illuminated and is flashing.
      Battery less than 10 %.

      Release method

      Replace the battery. Refer to Operation Manual "3.10 Installing foot control batteries"

      Error indication

      and are flashing alternately with beep sound.

      Cause of error

      The ON/OFF button of the foot control is depressed even though the chamber is not attached.

      Release method

      Attach the chamber to the control unit then operate the powder mode.

      Error indication

      and / are displayed, and 0 and are flashing with beep sound.

      Cause of error

      The ON/OFF button of the foot control is depressed even though the irrigation level is "0".

      Release method

      Rotate the water volume adjustment dial and adjust the irrigation level (1 - 10).

      Error indication

      G/E/P and are flashing alternately with beep sound.

      Cause of error

      Keep pressing the ON/OFF button when exiting the sleep mode.

      Release method

      Release the ON/OFF button.

      Error indication

      and /// are flashing alternately with beep sound.

      Cause of error

      Keep pressing the ON/OFF button when exiting the sleep mode.

      Release method

      Release the ON/OFF button.

    • A

      ERROR 0

      Description of error
      System error
      Check/Remedy
      Turn off the control unit, and then turn it on again. If the error still persists, contact your authorized NSK dealer.
      ERROR 1

      Description of error
      24 V voltage abnormality
      Check/Remedy
      Turn off the control unit, and then turn it on again. If the error still persists, contact your authorized NSK dealer.
      ERROR 2

      Description of error
      12 V voltage abnormality
      Check/Remedy
      Turn off the control unit, and then turn it on again. If the error still persists, contact your authorized NSK dealer.
      ERROR 3

      Description of error
      Pump current abnormality
      Check/Remedy
      Check the connection of the irrigation pump. If the error still persists, contact your authorized NSK dealer.
      ERROR 4

      Description of error
      Air pressure abnormality
      Check/Remedy
      Check the incoming air pressure. If the error still persists, contact your authorized NSK dealer.
      ERROR 5

      Description of error
      Heater voltage abnormality
      Check/Remedy
      Turn off the control unit, and then turn it on again. If the error still persists, contact your authorized NSK dealer.
      ERROR 6

      Description of error
      Heater current abnormality
      Check/Remedy
      Turn off the control unit, and then turn it on again. If the error still persists, contact your authorized NSK dealer.
      ERROR 7

      Description of error
      Heater temperature abnormality
      Check/Remedy
      Check the water circuit, and then supply the control unit with water. Cool down the control unit, and then operate it again. If the error still persists, contact your authorized NSK dealer.
      ERROR 8

      Description of error
      Foot control communication abnormality
      Check/Remedy
      Check the connection of the foot control. If the error still persists, contact your authorized NSK dealer.
      ERROR 9

      Description of error
      Ultrasonic handpiece abnormality
      Check/Remedy
      Check the connection of the ultrasonic handpiece. If the error still persists, contact your authorized NSK dealer.
      ERROR 10

      Description of error
      Air proportional control valve current abnormality
      Check/Remedy
      Turn off the control unit, and then turn it on again. If the error still persists, contact your authorized NSK dealer.
      ERROR 12

      Description of error
      Touch sensor communication abnormality
      Check/Remedy
      Turn off the control unit, and then turn it on again. If the error still persists, contact your authorized NSK dealer.
      ERROR 13

      Description of error
      Ultrasonic unit self-check error
      Check/Remedy
      Turn off the control unit, and then turn it on again. If the error still persists, contact your authorized NSK dealer.
      ERROR 14

      Description of error
      Ultrasonic unit system error
      Check/Remedy
      Turn off the control unit, and then turn it on again. If the error still persists, contact your authorized NSK dealer.
    • A

      Cause1
      The plug is improperly inserted to power source.
      Remedy1
      Check to make sure that the plug is properly connected.
      Cause2
      Malfunction of the power switch.
      Remedy2
      Contact your authorized NSK dealer*.
      Cause3
      The fuse has blown.
      Remedy3
      Contact your authorized NSK dealer*.

      * Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.

    • A

      Cause1
      Power was turned on while the foot control was depressed.
      Remedy1
      Release the foot control and turn the power switch on.
      Cause2
      Abnormal heating of the control unit.
      Remedy2
      Stop the operation and wait until the control unit cools down.
    • A

      Cause1
      Calibration is slightly misaligned due to static electricity, exogenous noise, etc.
      Remedy1
      Turn the power switch off and then on.
      Cause2
      The key was not touched at the center.
      Remedy2
      Touch the center of the key.
      Cause3
      Malfunction of the control unit.
      Remedy3
      Contact your authorized NSK dealer*.

      * Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.

    • A

      Cause1
      Water volume adjustment dial has been damaged.
      Remedy1
      Contact your authorized NSK dealer*.
      Cause2
      Damage or wear of an O-ring.
      Remedy2
      Replace the O-ring with a new one. Refer to Operation Manual "6.1.5 Chamber connector"
      Cause3
      The water circuit has been clogged with foreign materials.
      Remedy3
      Perform auto-cleaning. If the symptom is still not resolved, contact your authorized NSK dealer.

      * Cannot be fixed at dental clinics. Be sure to contact your authorized NSK deale

    • A

      Cause1
      Malfunction of the control unit.
      Remedy1
      Contact your authorized NSK dealer*.
      Cause2
      Water temperature setting is "OFF".
      Remedy2
      Set warm water to "ON". Refer to Operation Manual "7.3.2 Setting warm water"

      * Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.

    • A

      Cause1
      Bottle lid is not tightened firmly.
      Remedy1
      Firmly tighten the bottle lid.
      Cause2
      Worn O-ring or gasket of the bottle lid.
      Remedy2
      Replace the O-ring or gasket with a new one.
      Cause3
      Irrigation bottle is not inserted all the way. (There is a gap between the irrigation bottle and the control unit.)
      Remedy3
      Insert the irrigation bottle all the way. Make sure that there is no gap between the irrigation bottle and the control unit.
    • A

      Cause

      The foot control could not be connected.

      Remedy

      Perform the following steps.
      1. When "PUSH FC" is displayed on the LCD, depress and hold the ON/OFF button of the foot control.
      If the screen does not change after five minutes, touch
      2. Replace the batteries with new ones.
      3. Depress and hold the ON/OFF button of the foot control. If the screen does not change after five minutes, pair the control unit and the foot control again. Refer to Operation Manual "7.3.1 Pairing/unpairing foot control".
      If the screen does not change after performing the steps 1 through 3, contact your authorized NSK dealer.

    • A

      Cause1
      The foot control has not been connected.
      Remedy1
      Perform the following steps. 1. Remove any obstructions between the control unit and the foot control (which may be blocking communication). 2. Turn off the power switch, and then turn it on again. 3. Replace the batteries of the foot control with new ones. 4. Pair the control unit and the foot control. Refer to Operation Manual "7.3.1 Pairing/unpairing foot control" 5. If the foot control cannot connect with the control unit after performing the steps 1 through 4, contact your authorized NSK dealer.*
      Cause2
      Poor response of foot control buttons due to long periods of inactivity.
      Remedy2
      Perform calibration of the foot control. Refer to Operation Manual "7.3.8 Foot control calibration"
      Cause3
      Malfunction of the control unit or foot control.
      Remedy3
      Perform the following steps. 1. Replace the batteries of the foot control with new ones. 2. Pair the control unit and the foot control. Refer to Operation Manual "7.3.1 Pairing/unpairing foot control" 3. Depress the ON/OFF button of the foot control to confirm that the product is operating. If the product does not operate after performing the steps 1 through 3, repairs are required. Contact your authorized NSK dealer*.

      * Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.

    • A

      Cause
      Malfunction of the control unit.
      Remedy
      Contact your authorized NSK dealer*.

      * Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.

    • A

      Cause
      Malfunction of the control unit.
      Remedy
      Contact your authorized NSK dealer*.

      * Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.

    • A

      Cause

      The foot control has not been connected.

      Remedy

      Perform the following steps.
      1. Touch to cancel the error.
      2. Remove any obstructions between the control unit and the foot control (which may be blocking communication).
      3. Wait for the foot control to be automatically connected.

      If it is not automatically connected, perform the following procedure:
      1. Replace the batteries of the foot control with new ones.
      2. Pair the control unit and the foot control. Refer to Operation Manual "7.3.1 Pairing/unpairing foot control"
      3. Depress the ON/OFF button of the foot control to confirm that the product is operating.
      If the error persists after performing the steps 1 through 3, repairs are required. Contact your authorized NSK dealer*.

      * Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.

    • A

      Cause

      The batteries of the foot control are running down.

      Remedy

      Perform the following steps.
      1. Depress the ON/OFF button of the foot control or touch to cancel the "BATTERY" indication.
      2. Replace the batteries of the foot control with new ones.

      If the "BATTERY" indication does not disappear after performing the steps 1 through 2, repairs are required.
      Contact your authorized NSK dealer*.

      * Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.

    • A

      icon A:

      Cause
      The sensor built inside the ultrasonic handpiece holder is soiled and mistakenly detects mounting of the ultrasonic handpiece.
      Remedy
      Wipe off soil from the sensor built inside the ultrasonic handpiece holder.
    • A

      Cause1
      The AC power cord plug is not connected to the wall socket, or electricity is not supplied to the wall socket.
      Remedy1
      Check the connection.
      Cause2
      The fuse has blown.
      Remedy2
      Contact your authorized NSK dealer*.
      Cause3
      The tip is not tightened firmly.
      Remedy3
      Re-tighten the tip with the tip wrench by rotating the tip wrench until it makes a click sound and cannot rotate any further. If the problem still persists, replace the tip wrench as it may be worn out.
      Cause4
      The tip is worn.
      Remedy4
      Replace the tip with a new one.
      Cause5
      The threads of the ultrasonic handpiece and tip are not clean.
      Remedy5
      Wipe them clean.
      Cause6
      The threads of the ultrasonic handpiece are worn out as a result of long term use.
      Remedy6
      Contact your authorized NSK dealer*.
      Cause7
      Power has not been correctly adjusted for the tip.
      Remedy7
      Adjust the power so that it does not exceed the set upper limit value mentioned in the supplied operation manual of the tip. Refer to Operataion Manual "1.2 Related manuals"
      Cause8
      The foot control has not been connected.
      Remedy8
      Pair the control unit and the foot control. Refer to Operation Manual "7.3.1 Pairing/unpairing foot control"
      Cause9
      Failure of the vibration device in the ultrasonic handpiece.
      Remedy9
      Contact your authorized NSK dealer*.
      Cause10
      Failure of internal components of the foot control.
      Remedy10
      Contact your authorized NSK dealer*.
      Cause11
      The sensor built into the ultrasonic handpiece holder is dirty and mistakenly detects mounting of the ultrasonic handpiece.
      Remedy11
      Wipe the sensor clean.

      * Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.

    • A

      Cause
      Power has not been correctly adjusted for the tip.
      Remedy
      Adjust the power so that it does not exceed the set upper limit value mentioned in the supplied operation manual of the tip. Refer to Operation Manual "1.2 Related manuals"
    • A

      Cause
      The tip is not tightened firmly.
      Remedy
      Re-tighten the tip with the tip wrench by rotating the tip wrench until it makes a click sound and cannot rotate any further. If the problem still persists, replace the tip wrench as it may be worn out.
      Cause2
      The tip is not tightened firmly.
      Remedy2
      Re-tighten the tip with the tip wrench by rotating the tip wrench until it makes a click sound and cannot rotate any further. If the problem still persists, replace the tip wrench as it may be worn out.
      Cause3
      Failure of the ultrasonic handpiece vibration device or control units internal components.
      Remedy3
      Contact your authorized NSK dealer*.

      * Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.

    • A

      Cause1
      Power has not been correctly adjusted for the tip.
      Remedy1
      Adjust the power so that it does not exceed the set upper limit value mentioned in the supplied operation manual of the tip. Refer to Operation Manual "1.2 Related manuals"
      Cause2
      The tip is not tightened firmly.
      Remedy2
      Re-tighten the tip with the tip wrench by rotating the tip wrench until it makes a click sound and cannot rotate any further. If the problem still persists, replace the tip wrench as it may be worn out.
      Cause3
      Failure of the ultrasonic handpiece vibration device or control units internal components.
      Remedy3
      Contact your authorized NSK dealer*.

      * Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.

    • A

      Cause1
      Power has not been correctly adjusted for the tip.
      Remedy1
      Re-tighten the tip with the tip wrench by rotating the tip wrench until it makes a click sound and cannot rotate any further. If the problem still persists, replace the tip wrench as it may be worn out.
      Cause2
      The tip is not tightened firmly.
      Remedy2
      Re-tighten the tip with the tip wrench by rotating the tip wrench until it makes a click sound and cannot rotate any further. If the problem still persists, replace the tip wrench as it may be worn out.
      Cause3
      The threads of the ultrasonic handpiece and tip are not clean.
      Remedy3
      Wipe them clean.
      Cause4
      The threads of the ultrasonic handpiece are worn out as a result of long term use.
      Remedy4
      Contact your authorized NSK dealer.
      Cause5
      The irrigation level is insufficient or irrigation has stopped.
      Remedy5
      Irrigate at an adequate volume.
      Cause6
      Failure of the ultrasonic handpiece vibration device or control units internal components.
      Remedy6
      Contact your authorized NSK dealer.
    • A

      Cause1
      The pump tube of the irrigation pump is twisted. (When using the bottle)
      Remedy1
      Straighten the twisted irrigation tube.
      Cause2
      It is time to replace the irrigation pump. (When using the bottle)
      Remedy2
      Replace the irrigation pump with a new one. Refer to Operation Manual "6.2 Replacing the pump"
    • A

      Cause1
      Water is not supplied to the control unit.
      Remedy1
      Check the water circuit so that water is supplied to the control unit. Supply water pressure: 0.1-0.3 MPa
      Cause2
      The irrigation level has not been adjusted.
      Remedy2
      Adjust the irrigation level by rotating the water volume adjustment knob. Refer to Operation Manual "When using tap water" in "3.13.1 Setting irrigation mode".
      Cause3
      Water supply is sometimes unstable when the irrigation level is insufficient.
      Remedy3
      This is not a malfunction. Adjust the irrigation level by rotating the water volume adjustment knob. Refer to Operation Manual "When using tap water" in "3.13.1 Setting irrigation mode".
      Cause4
      The water filter is clogged with foreign substances.
      Remedy4
      Replace the water filter with a new one. Refer to Operation Manual "6.3 Replacing the water filter"
    • A

      Cause
      The water hose is not inserted into the water hose connector correctly.
      Remedy
      Insert the water hose into the water hose connector fully and firmly. Refer to Operation Manual "3.3 Connecting water hose"
    • A

      Cause1
      The O-ring of the connection part between the ultrasonic handpiece and handpiece cord is damaged or worn.
      Remedy1
      Replace the O-ring with a new one. Refer to Operation Manual "6.1.1 Handpiece cord"
      Cause2
      Failure in the control unit or handpiece cord.
      Remedy2
      Contact your authorized NSK dealer*.
      Cause3
      The glass rod of the ultrasonic handpiece is damaged as a result of a long term use.
      Remedy3
      Contact your authorized NSK dealer*.

      * Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.

    • A

      icon B:   icon C:

      Cause1
      The sensor built inside the powder handpiece holder is dirty and mistakenly detects mounting of the powder handpiece.
      Remedy1
      Wipe off soil from the sensor built inside the powder handpiece holder.
      Cause2
      The chamber is not attached securely.
      Remedy2
      Attach the chamber securely until you hear a click sound. Refer to Operation Manual "3.9 Attaching and removing chamber"
    • A

      icon D:   icon E:

      Cause
      Failure in the control unit (broken wire) Damage to the sensor built into the chamber.
      Remedy
      Contact your authorized NSK dealer*.

      * Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.

    • A

      Cause1
      Nozzle is deformed or damaged.
      Remedy1
      Replace the nozzle with a new one. Refer to Operation Manual "4.2.3 Operation" - "■ Attaching nozzle".
      Cause2
      The inside of the nozzle tip is clogged with powder.
      Remedy2
      Replace the nozzle tip with a new one. Refer to Operation Manual "4.2.3.2 Using in perio mode" - "■ Using nozzle tip".
      Cause3
      The inside of the powder handpiece is clogged with powder.
      Remedy3
      Perform the following steps. 1. Clean the nozzle base with the supplied cleaning wire (S). 2. Clean the nozzle end with the supplied cleaning file. 3. Clean the inside of the powder handpiece with the cleaning wire (L). If powder and air do not flow after the cleaning described above, perform ultrasonic cleaning with warm water. Refer to Operation Manual "5.8 Ultrasonic cleaning (weekly)"
      Cause4
      The inside of the chamber is clogged with powder.
      Remedy4
      Remove powder from the chamber. Refer to Operation Manual "5.5.1 Removing powder"
      Cause5
      Powder has partially solidified with water, etc., in the air passage.
      Remedy5
      Remove powder from the air passage. Refer to Operation Manual "5.2.1 Ultrasonic"
      Cause6
      Powder container has been left open exposing powder to the air.
      Remedy6
      Dispose of the powder and use new powder
      Cause7
      Water is mixed in supply air.
      Remedy7
      Open the filter in the air supply circuit to drain the compressor, and check that water has not accumulated inside the air filter and compressor drain.
      Cause8
      The element has been clogged.
      Remedy8
      Contact your authorized NSK dealer*.

      * Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.

    • A

      Cause
      The O-ring is scratched.
      Remedy
      Perform the following steps. 1. Replace the O-ring of the handpiece side connector of the handpiece hose. 2. Apply a drop of the supplied lubrication oil and rub it in with your finger. Refer to Operation Manual "6.1.2 Handpiece hose (Prophy-Hose2/Perio-Hose2)"
    • A

      Cause
      The gasket has been worn or damaged.
      Remedy
      Contact your authorized NSK dealer*.

      * Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.

    • A

      Cause
      Powder adheres to the threads of the chamber and/or the chamber lid.
      Remedy
      Clean the threads of the chamber lid. Refer to Operation Manual "5.5.3 Cleaning chamber lid"

    Varios Combi Pro

    • A

      Cause1
      Power was turned on while the Foot Control was depressed.
      Remedy1
      Release the Foot Control and turn the POWER switch to ON.
      Cause2
      Abnormal heating of the Control Unit.
      Remedy2
      Stop operation and leave until Control Unit cools.
    • A

      Cause1
      Calibration is slightly misaligned due to static electricity, exogenous noise, etc.
      Remedy1
      Turn the POWER switch to OFF and then to ON.
      Cause2
      Key was not touched in the center.
      Remedy2
      Touch the center of the key.
    • A

      Cause1
      Damage or wear of the O-Ring.
      Remedy1
      Follow the procedure in “10-1-5 Chamber Base Connector” to replace the O-Ring.
      Cause2
      Foreign material has become clogged in the water line.
      Remedy2
      Perform Auto-Cleaning. If the symptom is still not resolved, contact your Authorized NSK Dealer.
    • A

      Cause
      The sensor built in to the Handpiece Holder is dirty and mistakenly detects mounting of the Handpiece.
      Remedy
      Wipe the sensor clean.
    • A

      Cause1
      The AC power cord plug is not connected to the wall socket or electricity is not supplied to the wall socket.
      Remedy1
      Check the connection.
      Cause2
      The Fuse is burned out.
      Remedy2
      Refer to "10-5 Replacing the fuse" and replace the fuse.
      Cause3
      The Tip is not tightened firmly.
      Remedy3
      Rotate the Tip Wrench until it makes a click sound and cannot rotate any further. If the problem still persists, replace the Tip Wrench as it may be worn out.
      Cause4
      Worn Tip.
      Remedy4
      Replace the Tip.
      Cause5
      The screw parts of the Handpiece and Tip are not clean.
      Remedy5
      Wipe them clean.
      Cause6
      The screw parts of the Handpiece are worn out as a result of a long time of use.
      Remedy6
      Contact your Authorized NSK Dealer.*
      Cause7
      Power has not been correctly adjusted for the Tip.
      Remedy7
      Adjust the power so it does not exceed the maximum power level mentioned in Tip-Power Guide or on the Tip case label.
      Cause8
      The Foot Control is disconnected.
      Remedy8
      Connect the Foot Control Cord lead correctly.
      Cause9
      Failure of vibration device in the Handpiece.
      Remedy9
      Contact your Authorized NSK Dealer.*
      Cause10
      Failure of internal components of the Foot Control.
      Remedy 10
      Contact your Authorized NSK Dealer.*
      Cause11
      The sensor built in to the Handpiece Holder is dirty and mistakenly detects mounting of the Handpiece.
      Remedy 11
      Wipe the sensor clean.

      * Cannot be fixed at clinics, etc.

    • A

      Cause
      Power has not been properly adjusted for the Tip.
      Remedy
      Adjust the power so it does not exceed the maximum power level mentioned in Tip-Power Guide or the Tip case label.
    • A

      Cause
      The Tip is not tightened firmly.
      Remedy
      Tighten the Tip until the Tip Wrench clicks.
    • A

      Cause1
      Power has not been properly adjusted for the Tip.
      Remedy1
      Adjust the power so it does not exceed the maximum power level mentioned in Tip-Power Guide or the Tip case label.
      Cause2
      The Tip is not tightened firmly.
      Remedy2
      Rotate the Tip Wrench until it makes a click sound and cannot rotate any further. If the problem still persists, replace the Tip Wrench as it may be worn out.
      Cause3
      Failure of vibration in the Handpiece or the Control Unit.
      Remedy3
      Contact your Authorized NSK Dealer.*

      * Cannot be fixed at clinics, etc.

    • A

      Cause1
      Power has not been properly adjusted for the Tip.
      Remedy1
      Adjust the power so it does not exceed the maximum power level mentioned in Tip-Power Guide or the Tip case label.
      Remedy2
      The Tip is not tightened firmly.
      Remedy2
      Rotate the Tip Wrench until it makes a click sound and cannot rotate any further. If the problem still persists, replace the Tip Wrench as it may be worn out.
      Cause3
      The screw parts of the Handpiece and Tip are not clean.
      Remedy3
      Wipe them clean.
      Cause4
      The screw parts of the Handpiece are worn out as a result of a long time of use.
      Remedy4
      Contact your Authorized NSK Dealer.*
      Cause5
      The Irrigation Volume is insufficient or irrigation has stopped.
      Remedy5
      Irrigate at an adequate volume.
      Cause6
      Vibration failure in the Handpiece or the Control Unit.
      Remedy6
      Contact your Authorized NSK Dealer.*

      * Cannot be fixed at clinics, etc.

    • A

      Cause1
      The Irrigation Tube is twisted.
      Remedy1
      Straighten the twisted Irrigation Tube.
      Cause2
      Time to replace the Irrigation Pump.
      Remedy2
      Replace with a new Irrigation Pump (Refer to 10-2 Changing the Irrigation Pump).
    • A

      Cause1
      Water is not supplied to the Control Unit.
      Remedy1
      Check the water circuit so that water is supplied to the Control Unit. Supply water pressure: 0.1 - 0.3MPa
      Cause2
      The Irrigation Volume is set to “0.”
      Remedy2
      Adjust to the adequate Irrigation Volume using the Irrigation Volume setting.
      Cause3
      Water supply may be unstable when the Irrigation Volume is insufficient.
      Remedy3
      This is not a malfunction. Increase the Irrigation Volume in the Irrigation Volume setting.
      Cause4
      Water Filter is clogged with foreign substances.
      Remedy4
      Replace the water filter with a new one.
    • A

      Cause1
      The Irrigation Tube is not connected correctly.
      Remedy1
      Firmly insert the Irrigation Tube all the way into the Irrigation Connector.
      Cause2
      O-Ring at the Handpiece Cord is worn or damaged.
      Remedy2
      Replace with a new O-ring by following the procedures in "10-1-1 Handpiece Cord."
      Cause3
      The water circuitry in the Control Unit is damaged.
      Remedy3
      Contact your Authorized NSK Dealer.*

      * Cannot be fixed at clinics, etc.

    • A

      Cause1
      The Handpiece is not connected to the Handpiece Cord correctly.
      Remedy1
      Firmly insert the Handpiece all the way into the Handpiece Cord.
      Cause2
      Failure in the Control Unit.
      Remedy2
      Contact your Authorized NSK Dealer.*

      * Cannot be fixed at clinics, etc.

    • A

      Cause1
      The sensor built in to the Handpiece Holder is dirty and mistakenly detects mounting of the Handpiece.
      Remedy1
      Wipe the sensor clean.
      Cause2
      Chamber is not attached securely.
      Remedy2
      Attach the Chamber securely until the confirmation sound is heard.
    • A

      Cause
      Failure in the Control Unit (disconnection).
      Remedy
      Contact your Authorized NSK Dealer.*

      * Cannot be fixed at clinics, etc.

    • A

      Cause1
      The inside of the Nozzle Tip is clogged with powder.
      Remedy1
      Replace the Nozzle Tip.
      Cause2
      The inside of the Handpiece is clogged with powder.
      Remedy2
      Clean the Nozzle base with the supplied Cleaning wire (small) and Nozzle end with the cleaning file. In addition, clean the inside of the Handpiece with the Cleaning wire (large). If powder and air do not flow after being cleaned as described above, perform Ultrasonic Cleaning with warm water. (Refer to “9–3 Ultrasonic cleaning (weekly).”)
      Cause3
      The inside of the Chamber is clogged with powder.
      Remedy3
      Remove powder from inside the Chamber by following the procedures in "9–1–5 Chamber and Handpiece Hose".
      Cause4
      Powder has partially solidified with water, etc., in the air passage.
      Remedy4
      Remove powder from inside the passage by following the procedures in "9–1–4–2 Auto-Cleaning."
      Cause5
      Powder container has been left open exposing powder to the air.
      Remedy5
      Dispose of the powder and use new powder.
      Cause6
      Water is mixed in supply air.
      Remedy6
      Open the filter in the air supply circuit to drain the compressor, and check that water has not accumulated inside the air filter and compressor drain.
      Cause7
      The VCP Element is clogged.
      Remedy7
      Contact your Authorized NSK Dealer.*

      * Cannot be fixed at clinics, etc.

    • A

      Cause
      The O-ring is chipped.
      Remedy
      Replace the O-ring in the Handpiece joint section, apply the supplied turbine oil and rub it in with your finger. (Refer to "10-1-2 Handpiece Hose (for Prophy use and Perio use).")
    • A

      Cause
      The button is locked
      Remedy
      Follow the procedure in “Releasing the Lock Button” in “9–1–5 Chamber and Handpiece Hose” to release the Lock. If the symptom is not resolved, contact your Authorized NSK Dealer.*

      * Cannot be fixed at clinics, etc.

    • A

      Cause
      Powder is adhered to the threads of the Chamber and/or the Chamber Cap.
      Remedy
      Clean the threads of the Chamber Cap (Refer to “Cleaning Chamber Cap” in “9-1-5-2 Cleaning and disinfection”).

    Varios2 Ultrasonic Handpiece

    • A

      Cause1
      The threads of the ultrasonic handpiece are worn out as a result of a long time of use.
      Remedy1
      Wipe them clean.
      Cause2
      The threads of the ultrasonic handpiece are worn out as a result of long term use.
      Remedy2
      Contact your authorized NSK dealer*.
      Cause3
      Failure of the vibration device in the ultrasonic handpiece.
      Remedy3
      Contact your authorized NSK dealer*.

      * Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.

    • A

      Cause1
      Power has not been correctly adjusted for the tip.
      Remedy1
      Adjust the power so that it does not exceed the set upper limit value mentioned in the supplied operation manual of the tip. Refer to Operation Manual "Related manuals"
      Cause2
      The tip is not tightened firmly.
      Remedy2
      Re-tighten the tip with the tip wrench by rotating the tip wrench until it makes a click sound and cannot rotate any further. If the problem still persists, replace the tip wrench as it may be worn out.
      Cause3
      Failure of the ultrasonic handpiece's vibration device or control unit's internal components.
      Remedy3
      Contact your authorized NSK dealer*.

      * Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.

    • A

      Cause1
      Power has not been correctly adjusted for the tip.
      Remedy1
      Adjust the power so that it does not exceed the set upper limit value mentioned in the supplied operation manual of the tip. Refer to Operation Manual "Related manuals"
      Cause2
      The tip is not tightened firmly.
      Remedy2
      Re-tighten the tip with the tip wrench by rotating the tip wrench until it makes a click sound and cannot rotate any further. If the problem still persists, replace the tip wrench as it may be worn out.
      Cause3
      The threads of the ultrasonic handpiece are worn out as a result of a long time of use.
      Remedy3
      Wipe them clean.
      Cause4
      The threads of the ultrasonic handpiece are worn out as a result of long term use.
      Remedy4
      Contact your authorized NSK dealer*.
      Cause5
      Failure of the ultrasonic handpiece's vibration device or control unit's internal components.
      Remedy5
      Contact your authorized NSK dealer*.

      * Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.

    • A

      Cause
      The ultrasonic handpiece has been dropped or deteriorated, or unspecified chemicals have been used.
      Remedy
      Contact your authorized NSK dealer*.

      * Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.

    Water Bottle Kit

    • A

      Cause1
      Clogged filter due to deterioration over time.
      Remedy1
      Air filter needs to be replaced. Contact your authorized NSK dealer.
      Cause2
      Bent or pinched water hose.
      Remedy2
      Refer to Operation Manual "3.2 Connecting hose" and attach the water hose properly.
    • A

      Cause1
      Worn O-ring or gasket of the bottle lid.
      Remedy1
      Replace the O-ring or gasket with a new one.
      Cause2
      Damaged bottle or bottle lid.
      Remedy2
      Replace the bottle or bottle lid with a new one.
    • A

      Cause1
      Damaged drive part inside the water bottle module.
      Remedy1
      Contact your authorized NSK dealer.
      Cause2
      Damaged air circuit part inside the water bottle module.
      Remedy2
      Contact your authorized NSK dealer.
      Cause3
      A supply air pressure outside of the specified value was used.
      Remedy3
      Confirm the supply pressure and adjust to the specified pressure. If air continues to leak after adjustment, contact your authorized NSK dealer.
    • A

      Cause1
      The O-ring is worn.
      Remedy1
      Replace the O-ring.
      Cause2
      Damage inside the water bottle module.
      Remedy2
      Contact your authorized NSK dealer.

    Varios 970

    • A

      Probable Cause

      The Front Panel does not light, even if the Power Switch is ON.

      Cause1
      The Power Cord or the Jack is disconnected.
      Solution1
      Correctly insert the Power Cord or the Jack.
      Cause2
      The Fuse is burned out.
      Solution2
      Contact dealer.*

      * Repairs cannot be made by the customer.

    • A

      Probable Cause

      The Tip does not generate vibration, in spite of depressing the Foot Control

      Cause1
      The Tip is not tightened firmly.
      Solution1
      Tighten the Tip until the Tip Wrench clicks.
      Cause2
      Worn Tip.
      Solution2
      Replace the Tip.
      Cause3
      Power has not been correctly adjusted for the Tip.
      Solution3
      Adjust the power on the Power Guide or Tip case label. Do not exceed.
      Cause4
      The Foot Control is disconnected.
      Solution4
      Connect the Foot Control correctly.
      Cause5
      Failure of vibrator in the handpiece.
      Solution5
      Contact dealer.*
      Cause6
      Failure of internal components of the Foot Control.
      Solution6
      Contact dealer.*

      * Repairs cannot be made by the customer.

    • A

      Cause
      Power has not been properly adjusted for the Tip.
      Solution
      Adjust the power level on the power guide, as indicated on the Tip case label. Do not exceed the value shown.
    • A

      Cause
      The Tip is not tightened firmly.
      Solution
      Tighten the Tip until the Tip Wrench clicks.
    • A

      Cause1
      Power has not been properly adjusted for the Tip.
      Solution1
      Adjust the power level on the power guide, as indicated on the Tip case label. Do not exceed the value shown.
      Cause2
      The Tip is not tightened firmly.
      Solution2
      Tighten the Tip until the Tip Wrench clicks.
      Cause3
      Failure of vibration in the handpiece or the Control Unit.
      Solution3
      Contact dealer.*

      * Repairs cannot be made by the customer.

    • A

      Cause1
      Power has not been properly adjusted for the Tip.
      Solution1
      Adjust the power level on the power guide, as indicated on the Tip case label. Do not exceed the value shown.
      Cause2
      The Tip is not tightened firmly.
      Solution2
      Tighten the Tip until the Tip Wrench clicks.
      Cause3
      Failure of vibration in the handpiece or the Control Unit.
      Solution3
      Contact dealer.*

      * Repairs cannot be made by the customer.

    • A

      Probable Cause1
      The Irrigation Pump is running.
      Cause1
      The tube twisted.
      Solution1
      Straighten the twisted Irrigation Tube.
      Probable Cause2
      The Irrigation Pump is stopping.
      Cause2
      Time to replace Irrigation Pump. (Approx. 500hours after used.)
      Solution2
      Replace with new Irrigation Pump (Refer to 11. (4) Changing the Irrigation Pump).
    • A

      Probable Cause
      The water does not reach to the Control Unit.
      Solution
      Check the water circuitry and supply to the Control Unit. Water pressure : 0.1-0.5MPa (1-5kgf/cm²)
    • A

      Probable Cause

      Check to see if water reaches the Control Unit.

      Cause1
      The Water Adjustment Knob is closed.
      Solution1
      Turn the Water Adjustment Knob and adjust to the appropriate volume.
      Cause2
      Disconnected Irrigation supply at low volume range. (less than 10ml/min.)
      Solution2
      No problem. Turn the Water Adjustment Knob and increase the Irrigation volume.
      Cause3
      The Water Filter is clogged.
      Solution3
      Replace with new Water Filter (Refer to 11. (6) Changing Water Filter (Option) ).
    • A

      Probable Cause1
      Water is leaking from the joint between the Irrigation Tube and the Irrigation Connector.
      Cause1
      The Irrigation Tube is not connected correctly.
      Solution1
      Firmly insert the Irrigation Tube into the Irrigation Connector inmost.
      Probable Cause2
      Water is leaking from the joint between the handpiece and the cord.
      Cause2
      O-Ring at the handpiece cord is worn or damaged.
      Solution2
      Replace with new O-Ring (Refer to 11 (3) Changing O-Ring •Handpiece Cord).
      Probable Cause3
      Water is leaking from the Control Unit.
      Cause3
      The water circuitry in the Control Unit is damaged.
      Solution3
      Contact dealer.*

      * Repairs cannot be made by the customer.

    • A

      Probable Cause1
      Tip oscillates, but Handpiece LED turns on and off.
      Cause1
      The handpiece is not connected into the Handpiece Cord correctly.
      Solution1
      Firmly insert the handpiece into the Handpiece Cord inmost.
      Probable Cause2
      Tip oscillates, but Handpiece LED does not turn on.
      Cause2
      Disconnection in the Handpiece Cord, or failure in the Control Unit.
      Solution2
      Contact dealer.*

      * Repairs cannot be made by the customer.

    • A

      Probable Cause1
      Beeping while power on.
      Cause1
      Depress Foot Control.
      Solution1
      Release the Foot Control.
      Probable Cause2
      Beeping while stopping vibration of Tips.
      Cause2
      Abnormal heating of the Control Unit.
      Solution2
      Stop the operation and leave until Control Unit becomes cool.

    Varios 170

    • A

      Item to Check

      The Tip does not generate vibration, in spite of depressing the Foot Control.

      Cause1
      The Tip is not tightened firmly.
      Solution1
      Tighten the Tip until the Tip Wrench clicks.
      Cause2
      Worn Tip.
      Solution2
      Replace the Tip.
      Cause3
      Power has not been correctly adjusted for the Tip.
      Solution3
      Adjust the power level on the Power Guide.Do not exceed the maximum indicated on the tip label.
      Cause4
      Failure of vibrator in the handpiece.
      Solution4
      Contact dealer.*

      * Repairs cannot be made by the customer.

    • A

      Cause
      Power has not been properly adjusted for the Tip.
      Solution
      Adjust the power level on the Power Guide.Do not exceed the maximum indicated on the tip label.
    • A

      Cause
      The Tip is not tightened firmly.
      Solution
      Tighten the Tip until the Tip Wrench clicks.
    • A

      Cause1
      Power has not been properly adjusted for the Tip.
      Solution1
      Adjust the power level on the Power Guide. Do not exceed the maximum indicated on the tip label.
      Cause2
      The Tip is not tightened firmly.
      Solution2
      Tighten the Tip until the Tip Wrench clicks.
      Cause3
      Failure of vibration in the handpiece or the Module.
      Solution3
      Contact dealer.*

      * Repairs cannot be made by the customer.

    • A

      Cause1
      Power has not been properly adjusted for the Tip.
      Solution1
      Adjust the power level on the Power Guide. Do not exceed the maximum indicated on the tip label.
      Cause2
      The Tip is not tightened firmly.
      Solution2
      Tighten the Tip until the Tip Wrench clicks.
      Cause3
      Failure of vibration in the handpiece or the Module.
      Solution3
      Contact dealer.*

      * Repairs cannot be made by the customer.

    • A

      Item to Check
      The water does not reach to the handpiece.
      Solution
      Check the water circuitry and supply to the Unit Chair.
    • A

      Item to Check

      Remove the handpiece from the handpiece cord. Water does not come out or only a limited amount.

      Cause1
      The water adjustment dial is closed.
      Solution1
      Turn the water adjustment dial and set it to an appropriate water level.
      Cause2
      The water filter is clogged.
      Solution2
      Contact dealer.*

      * Repairs cannot be made by the customer.

    • A

      Item to Check

      No / Weak water from the handpiece.

      Cause
      There may be exogenous material in the hole.
      Solution
      Air it with the appliances such as syringes from the rear of the handpiece. If it is still unsolved, contact dealer.
    • A

      Item to Check

      Water is leaking from the joint between the handpiece and the cord.

      Cause
      O-Ring at the handpiece cord is worn or damaged.
      Solution
      Apply air from an appliance such as a 3in1 syringe to the rear of the handpiece. If it is still unsolved, contact dealer.
    • A

      Item to Check1
      Tip oscillates, but Handpiece LED turns on and off.
      Cause1
      The handpiece is not connected into the Handpiece Cord correctly.
      Solution1
      Firmly insert the handpiece into the Handpiece Cord.
      Item to Check2
      Tip oscillates, but Handpiece LED does not turn on.
      Cause2
      Disconnection in the Handpiece Cord, or failure in Module.
      Solution2
      Contact dealer.*

      * Repairs cannot be made by the customer.

    Chirurgia

    VarioSurg 4

    • A

      Cause of error
      Self-check error.
      Action
      Request repair.
    • A

      Cause of error1
      Tip is overloaded.
      Action1
      Be careful not to overload the tip. In the event of a situation where the tip is stuck in situe, step on the main pedal on the foot control and slowly pull out the tip while oscillating it (do not force it out).
      Cause of error2
      Tip not mounted, tip not tightened enough.
      Action2
      Install the tip and tighten it with the tip replacement wrench until it "clicks" and idles.
      Cause of error3
      Handpiece cord is not securely connected. Disconnection in the handpiece cord.
      Action3
      Check the handpiece cord connection. If this does not resolve it, contact an authorized NSK dealer.
    • A

      Cause of error
      Insufficient contact of the power cord. Failure of the power supply circuit.
      Action
      Securely re-connect the power cord. If the error cannot be eliminated, request repair.
    • A

      Cause of error
      Overheating by extended use under heavy load. Use immediately after the control unit is placed at high temperature (such as inside a car under the blazing sun or in a locker exposed to direct sunlight).
      Action
      Allow it to cool down before use. To allow heat to be sufficiently radiated, the periphery of the control unit should be well ventilated wherever possible. If the error cannot be eliminated, request repair.
    • A

      Cause of error
      Irrigation tube caught on pump roller. Irrigation pump failure.
      Action
      Check the irrigation tube. If the error cannot be eliminated, request repair.
    • A

      Cause of error
      Failure of Bluetooth communication circuit (LED handpiece system) inside the control unit.
      Action
      Request repair.
    • A

      Cause of error
      Failure of sensor in wireless foot control. Detection of disconnection of Bluetooth communication during ultrasonic vibration.
      Action
      Request repair.
    • A

      Cause of error
      Handpiece or tip not connected.
      Action
      Check that the handpiece cord is securely connected or that the tip is securely connected using a tip wrench. If not resolved, contact an authorized NSK dealer.
    • A

      Cause of error
      Abnormal control panel screen
      Action
      Request repair.
    • A

      Cause of error
      Failure of the Bluetooth communication circuit (link system with Surgic Pro2) inside the control unit.
      Action
      Request repair.
    • A

      Cause of error
      Failure of wired foot control.
      Action
      Make sure the connector of the wired foot control is securely connected. If not resolved, contact an authorized NSK dealer.
    • A

      Cause of problem1
      Plug improperly inserted to power source
      Action1
      Check to make sure the plug is in.
      Cause of problem2
      Malfunction of power switch
      Action2
      Request repair.
    • A

      Cause of problem1
      Power turned on while pressing the foot control
      Action1
      Remove foot from foot control.
      Cause of problem2
      Foot control malfunction
      Action2
      If the problem is solved by removing the batteries, it is a foot control malfunction. If not, it is a control unit malfunction. Please contact an authorized NSK dealer.
      Cause of problem3
      Control unit malfunction
      Action3
      If the problem is solved by removing the batteries, it is a foot control malfunction. If not, it is a control unit malfunction. Please contact an authorized NSK dealer.
    • A

      Cause of problem

      Wireless foot control is not connected.

      Action

      Keep pressing the main pedal of the wireless foot control while “PUSH FC” is displayed on the LCD.
      If the screen does not change after 5 minutes, press the button below.

      The normal screen is displayed.
      Replace the wireless foot control batteries to check if it can be connected.
      => “3.2 Installing batteries (Only for wireless foot control)”.
      If it cannot be connected after replacing the batteries, pairing the control unit and wireless foot control.
      => “7.2 Pairing with wireless foot control”.
      If this does not resolve the issue, please contact an authorized NSK dealer.

    • A

      Cause of problem1
      Wireless foot control is not connected.
      Action1
      Perform the following after confirming that nothing is blocking the signal between the wireless foot control and control unit. ・Turn the main power switch OFF, and ON again to connect to the control unit. ・Replace the wireless foot control batteries with new ones. => “3.2 Installing batteries (Only for wireless foot control)”. ・Perform pairing the control unit and wireless foot control. => “7.2 Pairing with wireless foot control”. If this does not resolve the issue, please contact an authorized NSK dealer.
      Cause of problem2
      Wireless foot control pedal and button response is poor due to not being used for a long period of time.
      Action2
      Calibrate the foot control. => “6.2 Foot control calibration”.

    Surgic Pro2

    • A

      Cause of error
      Self-check error.
      Action
      Request repair.
    • A

      Cause of error
      Extended use under heavy load. Short circuit in the motor. Motor cord (power line) failure.
      Action
      Electrical contact may be insufficient. Securely re-connect the motor cord. If the error cannot be eliminated, request repair.
    • A

      Cause of error
      Insufficient contact of the power cord. Failure of the power supply circuit.
      Action
      Securely re-connect the power cord. If the error cannot be eliminated, request repair.
    • A

      Cause of error
      Motor sensor failure. Unplugged motor cord. Motor cord (signal line) failure. Ingress of water into the motor.
      Action
      Electrical contact may be insufficient. Securely re-connect the motor cord. If the error cannot be eliminated, request repair. (Make sure to attach a protection plug when cleaning in a washer-disinfector.)
    • A

      Cause of error
      Overheating by extended use under heavy load. Use immediately after the control unit is placed at high temperature (such as inside a car under the blazing sun or in a locker exposed to direct sunlight).
      Action
      Allow it to cool down before use. To allow heat to be sufficiently radiated, the periphery of the control unit should be well-ventilated wherever possible. If the error cannot be eliminated, request repair.
    • A

      Cause of error
      Abnormal voltage generated in the start / stop switch circuit. Failure in the start / stop switch circuit.
      Action
      When rotation and stop are repeated in a short period of time, a circuit to suppress the sudden acceleration at the start of rotation may be activated. Wait for a few seconds before use. If the error cannot be eliminated, request repair.
    • A

      Cause of error
      Bur or drill attachment failure. Handpiece failure. Motor failure.
      Action
      The chuck may still be open, or not sufficiently closed. Securely close the chuck. If the error cannot be eliminated, request repair.
    • A

      Cause of error
      Incorrectly fitted irrigation tubing caught in the pump roller. Irrigation pump failure.
      Action
      Check the irrigation tube. If the error cannot be eliminated, request repair.
    • A

      Cause of error
      Failure of the Bluetooth communication circuit.
      Action
      Request repair.
    • A

      Cause of error
      Failure of the foot control sensor (Hall IC). Self-check error.
      Action
      Request repair.
    • A

      Cause of error
      Adhesion of the saline solution to the motor LED. Failure of LED circuit.
      Action
      Remove the saline solution completely. If the error cannot be eliminated, request repair.
    • A

      Cause of error
      Failure of the LCD display driver. Failure of the electrostatic sheet.
      Action
      Request repair.
    • A

      Cause of error
      Failure of the link cable. Failure of the Link Module circuit. Failure of Bluetooth communication circuit in the Link Module.
      Action
      Turn off the VarioSurg 3 and re-connect the Link Module securely. If the error cannot be eliminated, request repair.
    • A

      Cause of error
      Affected by electromagnetic interference emitted by electronic equipment. Failure of Bluetooth communication circuit in Osseo 100+.
      Action
      Keep sufficient distance between the tip of Osseo 100+ and electronic equipment. If the error cannot be eliminated, request repair.
    • A

      Cause of error
      OS of external terminal is not supported. Failure of the external terminal.
      Action
      Check the supported OS in the application. Also, reinstall the application. If the error cannot be eliminated, request repair.
    • A

      Cause of Problem1
      Plug improperly inserted to power source
      Action1
      Check to make sure the plug is in.
      Cause of Problem2
      Malfunction of power switch
      Action2
      Request repair.
    • A

      Cause of Problem1
      Power turned on while pressing the foot control
      Action1
      Remove foot from foot control.
      Cause of Problem2
      Foot control malfunction
      Action2
      If the problem is solved by removing the batteries, it is a foot control malfunction. If not, it is a control unit malfunction. Please contact an authorized NSK dealer.
      Cause of Problem3
      Control unit malfunction
      Action3
      If the problem is solved by removing the batteries, it is a foot control malfunction. If not, it is a control unit malfunction. Please contact an authorized NSK dealer.
    • A

      Cause of Problem
      Foot control is not connected.
      Action
      Keep pressing the speed control pedal of the foot control while “PUSH FC” is displayed on the LCD display. If the screen does not change after 5 minutes, press and hold the Cancel key. The normal screen is displayed. Refer to “3.2 Installing foot control batteries” and replace the foot control batteries to check if it can be connected. If it cannot be connected after replacing the batteries, Refer to “7.1.1 Pairing with foot control” for pairing. If this does not resolve the issue, please contact an authorized NSK dealer.
    • A

      Cause of Problem1
      Foot control is not connected.
      Action1
      Perform the following after confirming that nothing is blocking the signal between the foot control and control unit. Turn the main unit power off and on again to connect to the control unit. Refer to “3.2 Installing foot control batteries” and replace the foot control batteries with new ones. Refer to “7.1.1 Pairing with foot control” and perform pairing. If this does not resolve the issue, please contact an authorized NSK dealer.
      Cause of Problem2
      Foot control pedal and button response is poor due to not being used for a long period of time.
      Action2
      Refer to “6.3 Foot control calibration” and calibrate the foot control.
      Cause of Problem3
      Malfunction of control unit or foot control
      Action3
      If the problem is not solved by checking foot control batteries or foot control connection status, system repair is required.
    • A

      Cause of Problem
      Foot control not connected
      Action
      Press the Cancel key to cancel the error. After confirming that nothing is blocking the signal between the control unit and the foot control, wait for a while to check if the foot control is connected automatically. If it is not connected automatically, perform the following procedures. Turn the control unit power off and on again to connect to the foot control. Refer to “3.2 Installing foot control batteries” and replace the foot control batteries with new ones. Refer to “7.1.1 Pairing with foot control” and perform pairing. If this does not resolve the issue, please contact an authorized NSK dealer.
    • A

      Cause of Problem
      Link Module not connected
      Action
      Press the Cancel key to cancel the error. After confirming that nothing is blocking the signal between the control unit and the Link Module, wait for a while to check if the Link Module is connected automatically. If it is not connected automatically, perform the following procedures. • Turn off and on both this product and VarioSurg 3 connected to this product . • Refer to the Link Module manual and try the pairing again. If this does not resolve the issue, please contact an authorized NSK dealer.
    • A

      Cause of Problem
      Osseo 100+ not connected
      Action
      Press the Cancel key to cancel the error. After confirming that nothing is blocking the signal between the control unit and Osseo 100+, wait for a while to check if Osseo 100+ is connected automatically. If it is not connected automatically, perform the following procedures. • Turn off and on both this product and Osseo 100+ connected to this product. • Charge the battery sufficiently with the Osseo 100+ charger and try the automatic connection again. • Refer to the Osseo 100+ manual and try the pairing again. If this does not resolve the issue, please contact an authorized NSK dealer.
    • A

      Cause of Problem
      Foot control battery is low
      Action
      Depress the speed control pedal of the foot control or press the Cancel key to cancel the error. Replace the batteries. If this does not resolve the issue, please contact an authorized NSK dealer.
    • A

      Cause of Problem
      Osseo 100+ battery is low
      Action
      Depress the speed control pedal of the foot control or press the Cancel key to cancel the error. Charge the battery sufficiently with the Osseo 100+ charger and try the automatic connection again. If this does not resolve the issue, please contact an authorized NSK dealer.

    Higiena i konserwacja

    iCare

    • A

      Cause1
      The power lead is not plugged in.
      Remedy1
      Check the connection.
      Cause2
      The power lead is not connected to the iCare.
      Remedy2
      Check the connection.
      Cause3
      The iCare is not switched ON.
      Remedy3
      Turn ON the black power switch (right side of iCare).
      Cause4
      The internal fuse has blown.
      Remedy4
      Replace the fuse.
    • A

      Cause1
      Oiling ports (1-4) are not switched on (the Connector LED is OFF).
      Remedy1
      Press one or more of the oiling ports.
      Cause2
      The door is OPEN before or during operation.
      Remedy2
      Close the door.
      Cause3
      The door is CLOSED before Chuck Lubrication.
      Remedy3
      Open the door before Chuck Lubrication.
      Cause4
      Oil chamber is empty (Error LED (Oil) lights up).
      Remedy4
      Fill with NSK iCare maintenance oil.
      Cause5
      Air is not supplied. Or the air pressure is low (Error LED (Air Pressure) lights up).
      Remedy5
      Check air supply and air pressure.
      Cause6
      The room temperature is too low.
      Remedy6
      The iCare will operate in temperatures of 10 - 40°C. Especially in winter, wait until the room temperature reaches the operating temperature.
    • A

      Cause
      The hole of the Chuck Lubricate Nozzle is clogged with debris.
      Remedy
      Remove dirt in the hole using a brush (Do not use a metal brush).
    Rozwiązywanie problemów | Wsparcie | NSK Poland