Rozwiązywanie problemów
Higiena jamy ustnej
Varios Combi Pro2
- AError indication
Cause of errorWireless communication disconnected during ultrasonic mode or powder mode.
Release methodTouch
Error indication
Cause of errorWireless communication disconnected during ultrasonic mode or powder mode.
Release methodDepress the ON/OFF button.
Error indication
"Battery" is displayed and
is flashing.Cause of error"BATTERY" is illuminated and
is flashing.
Battery less than 10 %.Release methodReplace the battery. Refer to Operation Manual "3.10 Installing foot control batteries"
Error indication
and
are flashing alternately with beep sound.Cause of errorThe ON/OFF button of the foot control is depressed even though the chamber is not attached.
Release methodAttach the chamber to the control unit then operate the powder mode.
Error indication
and
/
are displayed, and 0 and
are flashing with beep sound.Cause of errorThe ON/OFF button of the foot control is depressed even though the irrigation level is "0".
Release methodRotate the water volume adjustment dial and adjust the irrigation level (1 - 10).
Error indicationG/E/P and
are flashing alternately with beep sound.Cause of errorKeep pressing the ON/OFF button when exiting the sleep mode.
Release methodRelease the ON/OFF button.
Error indication
and
/
/
/
are flashing alternately with beep sound.Cause of errorKeep pressing the ON/OFF button when exiting the sleep mode.
Release methodRelease the ON/OFF button.
- AERROR 0Description of errorSystem errorCheck/RemedyTurn off the control unit, and then turn it on again. If the error still persists, contact your authorized NSK dealer.ERROR 1Description of error24 V voltage abnormalityCheck/RemedyTurn off the control unit, and then turn it on again. If the error still persists, contact your authorized NSK dealer.ERROR 2Description of error12 V voltage abnormalityCheck/RemedyTurn off the control unit, and then turn it on again. If the error still persists, contact your authorized NSK dealer.ERROR 3Description of errorPump current abnormalityCheck/RemedyCheck the connection of the irrigation pump. If the error still persists, contact your authorized NSK dealer.ERROR 4Description of errorAir pressure abnormalityCheck/RemedyCheck the incoming air pressure. If the error still persists, contact your authorized NSK dealer.ERROR 5Description of errorHeater voltage abnormalityCheck/RemedyTurn off the control unit, and then turn it on again. If the error still persists, contact your authorized NSK dealer.ERROR 6Description of errorHeater current abnormalityCheck/RemedyTurn off the control unit, and then turn it on again. If the error still persists, contact your authorized NSK dealer.ERROR 7Description of errorHeater temperature abnormalityCheck/RemedyCheck the water circuit, and then supply the control unit with water. Cool down the control unit, and then operate it again. If the error still persists, contact your authorized NSK dealer.ERROR 8Description of errorFoot control communication abnormalityCheck/RemedyCheck the connection of the foot control. If the error still persists, contact your authorized NSK dealer.ERROR 9Description of errorUltrasonic handpiece abnormalityCheck/RemedyCheck the connection of the ultrasonic handpiece. If the error still persists, contact your authorized NSK dealer.ERROR 10Description of errorAir proportional control valve current abnormalityCheck/RemedyTurn off the control unit, and then turn it on again. If the error still persists, contact your authorized NSK dealer.ERROR 12Description of errorTouch sensor communication abnormalityCheck/RemedyTurn off the control unit, and then turn it on again. If the error still persists, contact your authorized NSK dealer.ERROR 13Description of errorUltrasonic unit self-check errorCheck/RemedyTurn off the control unit, and then turn it on again. If the error still persists, contact your authorized NSK dealer.ERROR 14Description of errorUltrasonic unit system errorCheck/RemedyTurn off the control unit, and then turn it on again. If the error still persists, contact your authorized NSK dealer.
- ACause1The plug is improperly inserted to power source.Remedy1Check to make sure that the plug is properly connected.Cause2Malfunction of the power switch.Remedy2Contact your authorized NSK dealer*.Cause3The fuse has blown.Remedy3Contact your authorized NSK dealer*.
* Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.
- ACause1Power was turned on while the foot control was depressed.Remedy1Release the foot control and turn the power switch on.Cause2Abnormal heating of the control unit.Remedy2Stop the operation and wait until the control unit cools down.
- ACause1Calibration is slightly misaligned due to static electricity, exogenous noise, etc.Remedy1Turn the power switch off and then on.Cause2The key was not touched at the center.Remedy2Touch the center of the key.Cause3Malfunction of the control unit.Remedy3Contact your authorized NSK dealer*.
* Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.
- ACause1Water volume adjustment dial has been damaged.Remedy1Contact your authorized NSK dealer*.Cause2Damage or wear of an O-ring.Remedy2Replace the O-ring with a new one. Refer to Operation Manual "6.1.5 Chamber connector"Cause3The water circuit has been clogged with foreign materials.Remedy3Perform auto-cleaning. If the symptom is still not resolved, contact your authorized NSK dealer.
* Cannot be fixed at dental clinics. Be sure to contact your authorized NSK deale
- ACause1Malfunction of the control unit.Remedy1Contact your authorized NSK dealer*.Cause2Water temperature setting is "OFF".Remedy2Set warm water to "ON". Refer to Operation Manual "7.3.2 Setting warm water"
* Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.
- ACause1Bottle lid is not tightened firmly.Remedy1Firmly tighten the bottle lid.Cause2Worn O-ring or gasket of the bottle lid.Remedy2Replace the O-ring or gasket with a new one.Cause3Irrigation bottle is not inserted all the way. (There is a gap between the irrigation bottle and the control unit.)Remedy3Insert the irrigation bottle all the way. Make sure that there is no gap between the irrigation bottle and the control unit.
- ACause
The foot control could not be connected.
RemedyPerform the following steps.
1. When "PUSH FC" is displayed on the LCD, depress and hold the ON/OFF button of the foot control.
If the screen does not change after five minutes, touch
2. Replace the batteries with new ones.
3. Depress and hold the ON/OFF button of the foot control. If the screen does not change after five minutes, pair the control unit and the foot control again. Refer to Operation Manual "7.3.1 Pairing/unpairing foot control".
If the screen does not change after performing the steps 1 through 3, contact your authorized NSK dealer. - ACause1The foot control has not been connected.Remedy1Perform the following steps. 1. Remove any obstructions between the control unit and the foot control (which may be blocking communication). 2. Turn off the power switch, and then turn it on again. 3. Replace the batteries of the foot control with new ones. 4. Pair the control unit and the foot control. Refer to Operation Manual "7.3.1 Pairing/unpairing foot control" 5. If the foot control cannot connect with the control unit after performing the steps 1 through 4, contact your authorized NSK dealer.*Cause2Poor response of foot control buttons due to long periods of inactivity.Remedy2Perform calibration of the foot control. Refer to Operation Manual "7.3.8 Foot control calibration"Cause3Malfunction of the control unit or foot control.Remedy3Perform the following steps. 1. Replace the batteries of the foot control with new ones. 2. Pair the control unit and the foot control. Refer to Operation Manual "7.3.1 Pairing/unpairing foot control" 3. Depress the ON/OFF button of the foot control to confirm that the product is operating. If the product does not operate after performing the steps 1 through 3, repairs are required. Contact your authorized NSK dealer*.
* Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.
- ACauseMalfunction of the control unit.RemedyContact your authorized NSK dealer*.
* Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.
- ACauseMalfunction of the control unit.RemedyContact your authorized NSK dealer*.
* Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.
- ACause
The foot control has not been connected.
RemedyPerform the following steps.
1. Touch
to cancel the error.
2. Remove any obstructions between the control unit and the foot control (which may be blocking communication).
3. Wait for the foot control to be automatically connected.
If it is not automatically connected, perform the following procedure:
1. Replace the batteries of the foot control with new ones.
2. Pair the control unit and the foot control. Refer to Operation Manual "7.3.1 Pairing/unpairing foot control"
3. Depress the ON/OFF button of the foot control to confirm that the product is operating.
If the error persists after performing the steps 1 through 3, repairs are required. Contact your authorized NSK dealer*.* Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.
- ACause
The batteries of the foot control are running down.
RemedyPerform the following steps.
1. Depress the ON/OFF button of the foot control or touch
to cancel the "BATTERY" indication.
2. Replace the batteries of the foot control with new ones.
If the "BATTERY" indication does not disappear after performing the steps 1 through 2, repairs are required.
Contact your authorized NSK dealer*.* Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.
- A
icon A:
CauseThe sensor built inside the ultrasonic handpiece holder is soiled and mistakenly detects mounting of the ultrasonic handpiece.RemedyWipe off soil from the sensor built inside the ultrasonic handpiece holder. - ACause1The AC power cord plug is not connected to the wall socket, or electricity is not supplied to the wall socket.Remedy1Check the connection.Cause2The fuse has blown.Remedy2Contact your authorized NSK dealer*.Cause3The tip is not tightened firmly.Remedy3Re-tighten the tip with the tip wrench by rotating the tip wrench until it makes a click sound and cannot rotate any further. If the problem still persists, replace the tip wrench as it may be worn out.Cause4The tip is worn.Remedy4Replace the tip with a new one.Cause5The threads of the ultrasonic handpiece and tip are not clean.Remedy5Wipe them clean.Cause6The threads of the ultrasonic handpiece are worn out as a result of long term use.Remedy6Contact your authorized NSK dealer*.Cause7Power has not been correctly adjusted for the tip.Remedy7Adjust the power so that it does not exceed the set upper limit value mentioned in the supplied operation manual of the tip. Refer to Operataion Manual "1.2 Related manuals"Cause8The foot control has not been connected.Remedy8Pair the control unit and the foot control. Refer to Operation Manual "7.3.1 Pairing/unpairing foot control"Cause9Failure of the vibration device in the ultrasonic handpiece.Remedy9Contact your authorized NSK dealer*.Cause10Failure of internal components of the foot control.Remedy10Contact your authorized NSK dealer*.Cause11The sensor built into the ultrasonic handpiece holder is dirty and mistakenly detects mounting of the ultrasonic handpiece.Remedy11Wipe the sensor clean.
* Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.
- ACausePower has not been correctly adjusted for the tip.RemedyAdjust the power so that it does not exceed the set upper limit value mentioned in the supplied operation manual of the tip. Refer to Operation Manual "1.2 Related manuals"
- ACauseThe tip is not tightened firmly.RemedyRe-tighten the tip with the tip wrench by rotating the tip wrench until it makes a click sound and cannot rotate any further. If the problem still persists, replace the tip wrench as it may be worn out.Cause2The tip is not tightened firmly.Remedy2Re-tighten the tip with the tip wrench by rotating the tip wrench until it makes a click sound and cannot rotate any further. If the problem still persists, replace the tip wrench as it may be worn out.Cause3Failure of the ultrasonic handpiece vibration device or control units internal components.Remedy3Contact your authorized NSK dealer*.
* Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.
- ACause1Power has not been correctly adjusted for the tip.Remedy1Adjust the power so that it does not exceed the set upper limit value mentioned in the supplied operation manual of the tip. Refer to Operation Manual "1.2 Related manuals"Cause2The tip is not tightened firmly.Remedy2Re-tighten the tip with the tip wrench by rotating the tip wrench until it makes a click sound and cannot rotate any further. If the problem still persists, replace the tip wrench as it may be worn out.Cause3Failure of the ultrasonic handpiece vibration device or control units internal components.Remedy3Contact your authorized NSK dealer*.
* Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.
- ACause1Power has not been correctly adjusted for the tip.Remedy1Re-tighten the tip with the tip wrench by rotating the tip wrench until it makes a click sound and cannot rotate any further. If the problem still persists, replace the tip wrench as it may be worn out.Cause2The tip is not tightened firmly.Remedy2Re-tighten the tip with the tip wrench by rotating the tip wrench until it makes a click sound and cannot rotate any further. If the problem still persists, replace the tip wrench as it may be worn out.Cause3The threads of the ultrasonic handpiece and tip are not clean.Remedy3Wipe them clean.Cause4The threads of the ultrasonic handpiece are worn out as a result of long term use.Remedy4Contact your authorized NSK dealer.Cause5The irrigation level is insufficient or irrigation has stopped.Remedy5Irrigate at an adequate volume.Cause6Failure of the ultrasonic handpiece vibration device or control units internal components.Remedy6Contact your authorized NSK dealer.
- ACause1The pump tube of the irrigation pump is twisted. (When using the bottle)Remedy1Straighten the twisted irrigation tube.Cause2It is time to replace the irrigation pump. (When using the bottle)Remedy2Replace the irrigation pump with a new one. Refer to Operation Manual "6.2 Replacing the pump"
- ACause1Water is not supplied to the control unit.Remedy1Check the water circuit so that water is supplied to the control unit. Supply water pressure: 0.1-0.3 MPaCause2The irrigation level has not been adjusted.Remedy2Adjust the irrigation level by rotating the water volume adjustment knob. Refer to Operation Manual "When using tap water" in "3.13.1 Setting irrigation mode".Cause3Water supply is sometimes unstable when the irrigation level is insufficient.Remedy3This is not a malfunction. Adjust the irrigation level by rotating the water volume adjustment knob. Refer to Operation Manual "When using tap water" in "3.13.1 Setting irrigation mode".Cause4The water filter is clogged with foreign substances.Remedy4Replace the water filter with a new one. Refer to Operation Manual "6.3 Replacing the water filter"
- ACauseThe water hose is not inserted into the water hose connector correctly.RemedyInsert the water hose into the water hose connector fully and firmly. Refer to Operation Manual "3.3 Connecting water hose"
- ACause1The O-ring of the connection part between the ultrasonic handpiece and handpiece cord is damaged or worn.Remedy1Replace the O-ring with a new one. Refer to Operation Manual "6.1.1 Handpiece cord"Cause2Failure in the control unit or handpiece cord.Remedy2Contact your authorized NSK dealer*.Cause3The glass rod of the ultrasonic handpiece is damaged as a result of a long term use.Remedy3Contact your authorized NSK dealer*.
* Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.
- A
icon B:
icon C:
Cause1The sensor built inside the powder handpiece holder is dirty and mistakenly detects mounting of the powder handpiece.Remedy1Wipe off soil from the sensor built inside the powder handpiece holder.Cause2The chamber is not attached securely.Remedy2Attach the chamber securely until you hear a click sound. Refer to Operation Manual "3.9 Attaching and removing chamber" - A
icon D:
icon E:
CauseFailure in the control unit (broken wire) Damage to the sensor built into the chamber.RemedyContact your authorized NSK dealer*.* Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.
- ACause1Nozzle is deformed or damaged.Remedy1Replace the nozzle with a new one. Refer to Operation Manual "4.2.3 Operation" - "■ Attaching nozzle".Cause2The inside of the nozzle tip is clogged with powder.Remedy2Replace the nozzle tip with a new one. Refer to Operation Manual "4.2.3.2 Using in perio mode" - "■ Using nozzle tip".Cause3The inside of the powder handpiece is clogged with powder.Remedy3Perform the following steps. 1. Clean the nozzle base with the supplied cleaning wire (S). 2. Clean the nozzle end with the supplied cleaning file. 3. Clean the inside of the powder handpiece with the cleaning wire (L). If powder and air do not flow after the cleaning described above, perform ultrasonic cleaning with warm water. Refer to Operation Manual "5.8 Ultrasonic cleaning (weekly)"Cause4The inside of the chamber is clogged with powder.Remedy4Remove powder from the chamber. Refer to Operation Manual "5.5.1 Removing powder"Cause5Powder has partially solidified with water, etc., in the air passage.Remedy5Remove powder from the air passage. Refer to Operation Manual "5.2.1 Ultrasonic"Cause6Powder container has been left open exposing powder to the air.Remedy6Dispose of the powder and use new powderCause7Water is mixed in supply air.Remedy7Open the filter in the air supply circuit to drain the compressor, and check that water has not accumulated inside the air filter and compressor drain.Cause8The element has been clogged.Remedy8Contact your authorized NSK dealer*.
* Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.
- ACauseThe O-ring is scratched.RemedyPerform the following steps. 1. Replace the O-ring of the handpiece side connector of the handpiece hose. 2. Apply a drop of the supplied lubrication oil and rub it in with your finger. Refer to Operation Manual "6.1.2 Handpiece hose (Prophy-Hose2/Perio-Hose2)"
- ACauseThe gasket has been worn or damaged.RemedyContact your authorized NSK dealer*.
* Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.
- ACausePowder adheres to the threads of the chamber and/or the chamber lid.RemedyClean the threads of the chamber lid. Refer to Operation Manual "5.5.3 Cleaning chamber lid"
Varios Combi Pro
- ACause1Power was turned on while the Foot Control was depressed.Remedy1Release the Foot Control and turn the POWER switch to ON.Cause2Abnormal heating of the Control Unit.Remedy2Stop operation and leave until Control Unit cools.
- ACause1Calibration is slightly misaligned due to static electricity, exogenous noise, etc.Remedy1Turn the POWER switch to OFF and then to ON.Cause2Key was not touched in the center.Remedy2Touch the center of the key.
- ACause1Damage or wear of the O-Ring.Remedy1Follow the procedure in “10-1-5 Chamber Base Connector” to replace the O-Ring.Cause2Foreign material has become clogged in the water line.Remedy2Perform Auto-Cleaning. If the symptom is still not resolved, contact your Authorized NSK Dealer.
- ACauseThe sensor built in to the Handpiece Holder is dirty and mistakenly detects mounting of the Handpiece.RemedyWipe the sensor clean.
- ACause1The AC power cord plug is not connected to the wall socket or electricity is not supplied to the wall socket.Remedy1Check the connection.Cause2The Fuse is burned out.Remedy2Refer to "10-5 Replacing the fuse" and replace the fuse.Cause3The Tip is not tightened firmly.Remedy3Rotate the Tip Wrench until it makes a click sound and cannot rotate any further. If the problem still persists, replace the Tip Wrench as it may be worn out.Cause4Worn Tip.Remedy4Replace the Tip.Cause5The screw parts of the Handpiece and Tip are not clean.Remedy5Wipe them clean.Cause6The screw parts of the Handpiece are worn out as a result of a long time of use.Remedy6Contact your Authorized NSK Dealer.*Cause7Power has not been correctly adjusted for the Tip.Remedy7Adjust the power so it does not exceed the maximum power level mentioned in Tip-Power Guide or on the Tip case label.Cause8The Foot Control is disconnected.Remedy8Connect the Foot Control Cord lead correctly.Cause9Failure of vibration device in the Handpiece.Remedy9Contact your Authorized NSK Dealer.*Cause10Failure of internal components of the Foot Control.Remedy 10Contact your Authorized NSK Dealer.*Cause11The sensor built in to the Handpiece Holder is dirty and mistakenly detects mounting of the Handpiece.Remedy 11Wipe the sensor clean.
* Cannot be fixed at clinics, etc.
- ACausePower has not been properly adjusted for the Tip.RemedyAdjust the power so it does not exceed the maximum power level mentioned in Tip-Power Guide or the Tip case label.
- ACauseThe Tip is not tightened firmly.RemedyTighten the Tip until the Tip Wrench clicks.
- ACause1Power has not been properly adjusted for the Tip.Remedy1Adjust the power so it does not exceed the maximum power level mentioned in Tip-Power Guide or the Tip case label.Cause2The Tip is not tightened firmly.Remedy2Rotate the Tip Wrench until it makes a click sound and cannot rotate any further. If the problem still persists, replace the Tip Wrench as it may be worn out.Cause3Failure of vibration in the Handpiece or the Control Unit.Remedy3Contact your Authorized NSK Dealer.*
* Cannot be fixed at clinics, etc.
- ACause1Power has not been properly adjusted for the Tip.Remedy1Adjust the power so it does not exceed the maximum power level mentioned in Tip-Power Guide or the Tip case label.Remedy2The Tip is not tightened firmly.Remedy2Rotate the Tip Wrench until it makes a click sound and cannot rotate any further. If the problem still persists, replace the Tip Wrench as it may be worn out.Cause3The screw parts of the Handpiece and Tip are not clean.Remedy3Wipe them clean.Cause4The screw parts of the Handpiece are worn out as a result of a long time of use.Remedy4Contact your Authorized NSK Dealer.*Cause5The Irrigation Volume is insufficient or irrigation has stopped.Remedy5Irrigate at an adequate volume.Cause6Vibration failure in the Handpiece or the Control Unit.Remedy6Contact your Authorized NSK Dealer.*
* Cannot be fixed at clinics, etc.
- ACause1The Irrigation Tube is twisted.Remedy1Straighten the twisted Irrigation Tube.Cause2Time to replace the Irrigation Pump.Remedy2Replace with a new Irrigation Pump (Refer to 10-2 Changing the Irrigation Pump).
- ACause1Water is not supplied to the Control Unit.Remedy1Check the water circuit so that water is supplied to the Control Unit. Supply water pressure: 0.1 - 0.3MPaCause2The Irrigation Volume is set to “0.”Remedy2Adjust to the adequate Irrigation Volume using the Irrigation Volume setting.Cause3Water supply may be unstable when the Irrigation Volume is insufficient.Remedy3This is not a malfunction. Increase the Irrigation Volume in the Irrigation Volume setting.Cause4Water Filter is clogged with foreign substances.Remedy4Replace the water filter with a new one.
- ACause1The Irrigation Tube is not connected correctly.Remedy1Firmly insert the Irrigation Tube all the way into the Irrigation Connector.Cause2O-Ring at the Handpiece Cord is worn or damaged.Remedy2Replace with a new O-ring by following the procedures in "10-1-1 Handpiece Cord."Cause3The water circuitry in the Control Unit is damaged.Remedy3Contact your Authorized NSK Dealer.*
* Cannot be fixed at clinics, etc.
- ACause1The Handpiece is not connected to the Handpiece Cord correctly.Remedy1Firmly insert the Handpiece all the way into the Handpiece Cord.Cause2Failure in the Control Unit.Remedy2Contact your Authorized NSK Dealer.*
* Cannot be fixed at clinics, etc.
- ACause1The sensor built in to the Handpiece Holder is dirty and mistakenly detects mounting of the Handpiece.Remedy1Wipe the sensor clean.Cause2Chamber is not attached securely.Remedy2Attach the Chamber securely until the confirmation sound is heard.
- ACauseFailure in the Control Unit (disconnection).RemedyContact your Authorized NSK Dealer.*
* Cannot be fixed at clinics, etc.
- ACause1The inside of the Nozzle Tip is clogged with powder.Remedy1Replace the Nozzle Tip.Cause2The inside of the Handpiece is clogged with powder.Remedy2Clean the Nozzle base with the supplied Cleaning wire (small) and Nozzle end with the cleaning file. In addition, clean the inside of the Handpiece with the Cleaning wire (large). If powder and air do not flow after being cleaned as described above, perform Ultrasonic Cleaning with warm water. (Refer to “9–3 Ultrasonic cleaning (weekly).”)Cause3The inside of the Chamber is clogged with powder.Remedy3Remove powder from inside the Chamber by following the procedures in "9–1–5 Chamber and Handpiece Hose".Cause4Powder has partially solidified with water, etc., in the air passage.Remedy4Remove powder from inside the passage by following the procedures in "9–1–4–2 Auto-Cleaning."Cause5Powder container has been left open exposing powder to the air.Remedy5Dispose of the powder and use new powder.Cause6Water is mixed in supply air.Remedy6Open the filter in the air supply circuit to drain the compressor, and check that water has not accumulated inside the air filter and compressor drain.Cause7The VCP Element is clogged.Remedy7Contact your Authorized NSK Dealer.*
* Cannot be fixed at clinics, etc.
- ACauseThe O-ring is chipped.RemedyReplace the O-ring in the Handpiece joint section, apply the supplied turbine oil and rub it in with your finger. (Refer to "10-1-2 Handpiece Hose (for Prophy use and Perio use).")
- ACauseThe button is lockedRemedyFollow the procedure in “Releasing the Lock Button” in “9–1–5 Chamber and Handpiece Hose” to release the Lock. If the symptom is not resolved, contact your Authorized NSK Dealer.*
* Cannot be fixed at clinics, etc.
- ACausePowder is adhered to the threads of the Chamber and/or the Chamber Cap.RemedyClean the threads of the Chamber Cap (Refer to “Cleaning Chamber Cap” in “9-1-5-2 Cleaning and disinfection”).
Varios2 Ultrasonic Handpiece
- ACause1The threads of the ultrasonic handpiece are worn out as a result of a long time of use.Remedy1Wipe them clean.Cause2The threads of the ultrasonic handpiece are worn out as a result of long term use.Remedy2Contact your authorized NSK dealer*.Cause3Failure of the vibration device in the ultrasonic handpiece.Remedy3Contact your authorized NSK dealer*.
* Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.
- ACause1Power has not been correctly adjusted for the tip.Remedy1Adjust the power so that it does not exceed the set upper limit value mentioned in the supplied operation manual of the tip. Refer to Operation Manual "Related manuals"Cause2The tip is not tightened firmly.Remedy2Re-tighten the tip with the tip wrench by rotating the tip wrench until it makes a click sound and cannot rotate any further. If the problem still persists, replace the tip wrench as it may be worn out.Cause3Failure of the ultrasonic handpiece's vibration device or control unit's internal components.Remedy3Contact your authorized NSK dealer*.
* Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.
- ACause1Power has not been correctly adjusted for the tip.Remedy1Adjust the power so that it does not exceed the set upper limit value mentioned in the supplied operation manual of the tip. Refer to Operation Manual "Related manuals"Cause2The tip is not tightened firmly.Remedy2Re-tighten the tip with the tip wrench by rotating the tip wrench until it makes a click sound and cannot rotate any further. If the problem still persists, replace the tip wrench as it may be worn out.Cause3The threads of the ultrasonic handpiece are worn out as a result of a long time of use.Remedy3Wipe them clean.Cause4The threads of the ultrasonic handpiece are worn out as a result of long term use.Remedy4Contact your authorized NSK dealer*.Cause5Failure of the ultrasonic handpiece's vibration device or control unit's internal components.Remedy5Contact your authorized NSK dealer*.
* Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.
- ACauseThe ultrasonic handpiece has been dropped or deteriorated, or unspecified chemicals have been used.RemedyContact your authorized NSK dealer*.
* Cannot be fixed at dental clinics. Be sure to contact your authorized NSK dealer.
Water Bottle Kit
- ACause1Clogged filter due to deterioration over time.Remedy1Air filter needs to be replaced. Contact your authorized NSK dealer.Cause2Bent or pinched water hose.Remedy2Refer to Operation Manual "3.2 Connecting hose" and attach the water hose properly.
- ACause1Worn O-ring or gasket of the bottle lid.Remedy1Replace the O-ring or gasket with a new one.Cause2Damaged bottle or bottle lid.Remedy2Replace the bottle or bottle lid with a new one.
- ACause1Damaged drive part inside the water bottle module.Remedy1Contact your authorized NSK dealer.Cause2Damaged air circuit part inside the water bottle module.Remedy2Contact your authorized NSK dealer.Cause3A supply air pressure outside of the specified value was used.Remedy3Confirm the supply pressure and adjust to the specified pressure. If air continues to leak after adjustment, contact your authorized NSK dealer.
- ACause1The O-ring is worn.Remedy1Replace the O-ring.Cause2Damage inside the water bottle module.Remedy2Contact your authorized NSK dealer.
Varios 970
- AProbable Cause
The Front Panel does not light, even if the Power Switch is ON.
Cause1The Power Cord or the Jack is disconnected.Solution1Correctly insert the Power Cord or the Jack.Cause2The Fuse is burned out.Solution2Contact dealer.** Repairs cannot be made by the customer.
- AProbable Cause
The Tip does not generate vibration, in spite of depressing the Foot Control
Cause1The Tip is not tightened firmly.Solution1Tighten the Tip until the Tip Wrench clicks.Cause2Worn Tip.Solution2Replace the Tip.Cause3Power has not been correctly adjusted for the Tip.Solution3Adjust the power on the Power Guide or Tip case label. Do not exceed.Cause4The Foot Control is disconnected.Solution4Connect the Foot Control correctly.Cause5Failure of vibrator in the handpiece.Solution5Contact dealer.*Cause6Failure of internal components of the Foot Control.Solution6Contact dealer.** Repairs cannot be made by the customer.
- ACausePower has not been properly adjusted for the Tip.SolutionAdjust the power level on the power guide, as indicated on the Tip case label. Do not exceed the value shown.
- ACauseThe Tip is not tightened firmly.SolutionTighten the Tip until the Tip Wrench clicks.
- ACause1Power has not been properly adjusted for the Tip.Solution1Adjust the power level on the power guide, as indicated on the Tip case label. Do not exceed the value shown.Cause2The Tip is not tightened firmly.Solution2Tighten the Tip until the Tip Wrench clicks.Cause3Failure of vibration in the handpiece or the Control Unit.Solution3Contact dealer.*
* Repairs cannot be made by the customer.
- ACause1Power has not been properly adjusted for the Tip.Solution1Adjust the power level on the power guide, as indicated on the Tip case label. Do not exceed the value shown.Cause2The Tip is not tightened firmly.Solution2Tighten the Tip until the Tip Wrench clicks.Cause3Failure of vibration in the handpiece or the Control Unit.Solution3Contact dealer.*
* Repairs cannot be made by the customer.
- AProbable Cause1The Irrigation Pump is running.Cause1The tube twisted.Solution1Straighten the twisted Irrigation Tube.Probable Cause2The Irrigation Pump is stopping.Cause2Time to replace Irrigation Pump. (Approx. 500hours after used.)Solution2Replace with new Irrigation Pump (Refer to 11. (4) Changing the Irrigation Pump).
- AProbable CauseThe water does not reach to the Control Unit.SolutionCheck the water circuitry and supply to the Control Unit. Water pressure : 0.1-0.5MPa (1-5kgf/cm²)
- AProbable Cause
Check to see if water reaches the Control Unit.
Cause1The Water Adjustment Knob is closed.Solution1Turn the Water Adjustment Knob and adjust to the appropriate volume.Cause2Disconnected Irrigation supply at low volume range. (less than 10ml/min.)Solution2No problem. Turn the Water Adjustment Knob and increase the Irrigation volume.Cause3The Water Filter is clogged.Solution3Replace with new Water Filter (Refer to 11. (6) Changing Water Filter (Option) ). - AProbable Cause1Water is leaking from the joint between the Irrigation Tube and the Irrigation Connector.Cause1The Irrigation Tube is not connected correctly.Solution1Firmly insert the Irrigation Tube into the Irrigation Connector inmost.Probable Cause2Water is leaking from the joint between the handpiece and the cord.Cause2O-Ring at the handpiece cord is worn or damaged.Solution2Replace with new O-Ring (Refer to 11 (3) Changing O-Ring •Handpiece Cord).Probable Cause3Water is leaking from the Control Unit.Cause3The water circuitry in the Control Unit is damaged.Solution3Contact dealer.*
* Repairs cannot be made by the customer.
- AProbable Cause1Tip oscillates, but Handpiece LED turns on and off.Cause1The handpiece is not connected into the Handpiece Cord correctly.Solution1Firmly insert the handpiece into the Handpiece Cord inmost.Probable Cause2Tip oscillates, but Handpiece LED does not turn on.Cause2Disconnection in the Handpiece Cord, or failure in the Control Unit.Solution2Contact dealer.*
* Repairs cannot be made by the customer.
- AProbable Cause1Beeping while power on.Cause1Depress Foot Control.Solution1Release the Foot Control.Probable Cause2Beeping while stopping vibration of Tips.Cause2Abnormal heating of the Control Unit.Solution2Stop the operation and leave until Control Unit becomes cool.
Varios 170
- AItem to Check
The Tip does not generate vibration, in spite of depressing the Foot Control.
Cause1The Tip is not tightened firmly.Solution1Tighten the Tip until the Tip Wrench clicks.Cause2Worn Tip.Solution2Replace the Tip.Cause3Power has not been correctly adjusted for the Tip.Solution3Adjust the power level on the Power Guide.Do not exceed the maximum indicated on the tip label.Cause4Failure of vibrator in the handpiece.Solution4Contact dealer.** Repairs cannot be made by the customer.
- ACausePower has not been properly adjusted for the Tip.SolutionAdjust the power level on the Power Guide.Do not exceed the maximum indicated on the tip label.
- ACauseThe Tip is not tightened firmly.SolutionTighten the Tip until the Tip Wrench clicks.
- ACause1Power has not been properly adjusted for the Tip.Solution1Adjust the power level on the Power Guide. Do not exceed the maximum indicated on the tip label.Cause2The Tip is not tightened firmly.Solution2Tighten the Tip until the Tip Wrench clicks.Cause3Failure of vibration in the handpiece or the Module.Solution3Contact dealer.*
* Repairs cannot be made by the customer.
- ACause1Power has not been properly adjusted for the Tip.Solution1Adjust the power level on the Power Guide. Do not exceed the maximum indicated on the tip label.Cause2The Tip is not tightened firmly.Solution2Tighten the Tip until the Tip Wrench clicks.Cause3Failure of vibration in the handpiece or the Module.Solution3Contact dealer.*
* Repairs cannot be made by the customer.
- AItem to CheckThe water does not reach to the handpiece.SolutionCheck the water circuitry and supply to the Unit Chair.
- AItem to Check
Remove the handpiece from the handpiece cord. Water does not come out or only a limited amount.
Cause1The water adjustment dial is closed.Solution1Turn the water adjustment dial and set it to an appropriate water level.Cause2The water filter is clogged.Solution2Contact dealer.** Repairs cannot be made by the customer.
- AItem to Check
No / Weak water from the handpiece.
CauseThere may be exogenous material in the hole.SolutionAir it with the appliances such as syringes from the rear of the handpiece. If it is still unsolved, contact dealer. - AItem to Check
Water is leaking from the joint between the handpiece and the cord.
CauseO-Ring at the handpiece cord is worn or damaged.SolutionApply air from an appliance such as a 3in1 syringe to the rear of the handpiece. If it is still unsolved, contact dealer. - AItem to Check1Tip oscillates, but Handpiece LED turns on and off.Cause1The handpiece is not connected into the Handpiece Cord correctly.Solution1Firmly insert the handpiece into the Handpiece Cord.Item to Check2Tip oscillates, but Handpiece LED does not turn on.Cause2Disconnection in the Handpiece Cord, or failure in Module.Solution2Contact dealer.*
* Repairs cannot be made by the customer.
Chirurgia
VarioSurg 4
- ACause of errorSelf-check error.ActionRequest repair.
- ACause of error1Tip is overloaded.Action1Be careful not to overload the tip. In the event of a situation where the tip is stuck in situe, step on the main pedal on the foot control and slowly pull out the tip while oscillating it (do not force it out).Cause of error2Tip not mounted, tip not tightened enough.Action2Install the tip and tighten it with the tip replacement wrench until it "clicks" and idles.Cause of error3Handpiece cord is not securely connected. Disconnection in the handpiece cord.Action3Check the handpiece cord connection. If this does not resolve it, contact an authorized NSK dealer.
- ACause of errorInsufficient contact of the power cord. Failure of the power supply circuit.ActionSecurely re-connect the power cord. If the error cannot be eliminated, request repair.
- ACause of errorOverheating by extended use under heavy load. Use immediately after the control unit is placed at high temperature (such as inside a car under the blazing sun or in a locker exposed to direct sunlight).ActionAllow it to cool down before use. To allow heat to be sufficiently radiated, the periphery of the control unit should be well ventilated wherever possible. If the error cannot be eliminated, request repair.
- ACause of errorIrrigation tube caught on pump roller. Irrigation pump failure.ActionCheck the irrigation tube. If the error cannot be eliminated, request repair.
- ACause of errorFailure of Bluetooth communication circuit (LED handpiece system) inside the control unit.ActionRequest repair.
- ACause of errorFailure of sensor in wireless foot control. Detection of disconnection of Bluetooth communication during ultrasonic vibration.ActionRequest repair.
- ACause of errorHandpiece or tip not connected.ActionCheck that the handpiece cord is securely connected or that the tip is securely connected using a tip wrench. If not resolved, contact an authorized NSK dealer.
- ACause of errorAbnormal control panel screenActionRequest repair.
- ACause of errorFailure of the Bluetooth communication circuit (link system with Surgic Pro2) inside the control unit.ActionRequest repair.
- ACause of errorFailure of wired foot control.ActionMake sure the connector of the wired foot control is securely connected. If not resolved, contact an authorized NSK dealer.
- ACause of problem1Plug improperly inserted to power sourceAction1Check to make sure the plug is in.Cause of problem2Malfunction of power switchAction2Request repair.
- ACause of problem1Power turned on while pressing the foot controlAction1Remove foot from foot control.Cause of problem2Foot control malfunctionAction2If the problem is solved by removing the batteries, it is a foot control malfunction. If not, it is a control unit malfunction. Please contact an authorized NSK dealer.Cause of problem3Control unit malfunctionAction3If the problem is solved by removing the batteries, it is a foot control malfunction. If not, it is a control unit malfunction. Please contact an authorized NSK dealer.
- ACause of problem
Wireless foot control is not connected.
ActionKeep pressing the main pedal of the wireless foot control while “PUSH FC” is displayed on the LCD.
If the screen does not change after 5 minutes, press the button below.
The normal screen is displayed.
Replace the wireless foot control batteries to check if it can be connected.
=> “3.2 Installing batteries (Only for wireless foot control)”.
If it cannot be connected after replacing the batteries, pairing the control unit and wireless foot control.
=> “7.2 Pairing with wireless foot control”.
If this does not resolve the issue, please contact an authorized NSK dealer. - ACause of problem1Wireless foot control is not connected.Action1Perform the following after confirming that nothing is blocking the signal between the wireless foot control and control unit. ・Turn the main power switch OFF, and ON again to connect to the control unit. ・Replace the wireless foot control batteries with new ones. => “3.2 Installing batteries (Only for wireless foot control)”. ・Perform pairing the control unit and wireless foot control. => “7.2 Pairing with wireless foot control”. If this does not resolve the issue, please contact an authorized NSK dealer.Cause of problem2Wireless foot control pedal and button response is poor due to not being used for a long period of time.Action2Calibrate the foot control. => “6.2 Foot control calibration”.
Surgic Pro2
- ACause of errorSelf-check error.ActionRequest repair.
- ACause of errorExtended use under heavy load. Short circuit in the motor. Motor cord (power line) failure.ActionElectrical contact may be insufficient. Securely re-connect the motor cord. If the error cannot be eliminated, request repair.
- ACause of errorInsufficient contact of the power cord. Failure of the power supply circuit.ActionSecurely re-connect the power cord. If the error cannot be eliminated, request repair.
- ACause of errorMotor sensor failure. Unplugged motor cord. Motor cord (signal line) failure. Ingress of water into the motor.ActionElectrical contact may be insufficient. Securely re-connect the motor cord. If the error cannot be eliminated, request repair. (Make sure to attach a protection plug when cleaning in a washer-disinfector.)
- ACause of errorOverheating by extended use under heavy load. Use immediately after the control unit is placed at high temperature (such as inside a car under the blazing sun or in a locker exposed to direct sunlight).ActionAllow it to cool down before use. To allow heat to be sufficiently radiated, the periphery of the control unit should be well-ventilated wherever possible. If the error cannot be eliminated, request repair.
- ACause of errorAbnormal voltage generated in the start / stop switch circuit. Failure in the start / stop switch circuit.ActionWhen rotation and stop are repeated in a short period of time, a circuit to suppress the sudden acceleration at the start of rotation may be activated. Wait for a few seconds before use. If the error cannot be eliminated, request repair.
- ACause of errorBur or drill attachment failure. Handpiece failure. Motor failure.ActionThe chuck may still be open, or not sufficiently closed. Securely close the chuck. If the error cannot be eliminated, request repair.
- ACause of errorIncorrectly fitted irrigation tubing caught in the pump roller. Irrigation pump failure.ActionCheck the irrigation tube. If the error cannot be eliminated, request repair.
- ACause of errorFailure of the Bluetooth communication circuit.ActionRequest repair.
- ACause of errorFailure of the foot control sensor (Hall IC). Self-check error.ActionRequest repair.
- ACause of errorAdhesion of the saline solution to the motor LED. Failure of LED circuit.ActionRemove the saline solution completely. If the error cannot be eliminated, request repair.
- ACause of errorFailure of the LCD display driver. Failure of the electrostatic sheet.ActionRequest repair.
- ACause of errorFailure of the link cable. Failure of the Link Module circuit. Failure of Bluetooth communication circuit in the Link Module.ActionTurn off the VarioSurg 3 and re-connect the Link Module securely. If the error cannot be eliminated, request repair.
- ACause of errorAffected by electromagnetic interference emitted by electronic equipment. Failure of Bluetooth communication circuit in Osseo 100+.ActionKeep sufficient distance between the tip of Osseo 100+ and electronic equipment. If the error cannot be eliminated, request repair.
- ACause of errorOS of external terminal is not supported. Failure of the external terminal.ActionCheck the supported OS in the application. Also, reinstall the application. If the error cannot be eliminated, request repair.
- ACause of Problem1Plug improperly inserted to power sourceAction1Check to make sure the plug is in.Cause of Problem2Malfunction of power switchAction2Request repair.
- ACause of Problem1Power turned on while pressing the foot controlAction1Remove foot from foot control.Cause of Problem2Foot control malfunctionAction2If the problem is solved by removing the batteries, it is a foot control malfunction. If not, it is a control unit malfunction. Please contact an authorized NSK dealer.Cause of Problem3Control unit malfunctionAction3If the problem is solved by removing the batteries, it is a foot control malfunction. If not, it is a control unit malfunction. Please contact an authorized NSK dealer.
- ACause of ProblemFoot control is not connected.ActionKeep pressing the speed control pedal of the foot control while “PUSH FC” is displayed on the LCD display. If the screen does not change after 5 minutes, press and hold the Cancel key. The normal screen is displayed. Refer to “3.2 Installing foot control batteries” and replace the foot control batteries to check if it can be connected. If it cannot be connected after replacing the batteries, Refer to “7.1.1 Pairing with foot control” for pairing. If this does not resolve the issue, please contact an authorized NSK dealer.
- ACause of Problem1Foot control is not connected.Action1Perform the following after confirming that nothing is blocking the signal between the foot control and control unit. Turn the main unit power off and on again to connect to the control unit. Refer to “3.2 Installing foot control batteries” and replace the foot control batteries with new ones. Refer to “7.1.1 Pairing with foot control” and perform pairing. If this does not resolve the issue, please contact an authorized NSK dealer.Cause of Problem2Foot control pedal and button response is poor due to not being used for a long period of time.Action2Refer to “6.3 Foot control calibration” and calibrate the foot control.Cause of Problem3Malfunction of control unit or foot controlAction3If the problem is not solved by checking foot control batteries or foot control connection status, system repair is required.
- ACause of ProblemFoot control not connectedActionPress the Cancel key to cancel the error. After confirming that nothing is blocking the signal between the control unit and the foot control, wait for a while to check if the foot control is connected automatically. If it is not connected automatically, perform the following procedures. Turn the control unit power off and on again to connect to the foot control. Refer to “3.2 Installing foot control batteries” and replace the foot control batteries with new ones. Refer to “7.1.1 Pairing with foot control” and perform pairing. If this does not resolve the issue, please contact an authorized NSK dealer.
- ACause of ProblemLink Module not connectedActionPress the Cancel key to cancel the error. After confirming that nothing is blocking the signal between the control unit and the Link Module, wait for a while to check if the Link Module is connected automatically. If it is not connected automatically, perform the following procedures. • Turn off and on both this product and VarioSurg 3 connected to this product . • Refer to the Link Module manual and try the pairing again. If this does not resolve the issue, please contact an authorized NSK dealer.
- ACause of ProblemOsseo 100+ not connectedActionPress the Cancel key to cancel the error. After confirming that nothing is blocking the signal between the control unit and Osseo 100+, wait for a while to check if Osseo 100+ is connected automatically. If it is not connected automatically, perform the following procedures. • Turn off and on both this product and Osseo 100+ connected to this product. • Charge the battery sufficiently with the Osseo 100+ charger and try the automatic connection again. • Refer to the Osseo 100+ manual and try the pairing again. If this does not resolve the issue, please contact an authorized NSK dealer.
- ACause of ProblemFoot control battery is lowActionDepress the speed control pedal of the foot control or press the Cancel key to cancel the error. Replace the batteries. If this does not resolve the issue, please contact an authorized NSK dealer.
- ACause of ProblemOsseo 100+ battery is lowActionDepress the speed control pedal of the foot control or press the Cancel key to cancel the error. Charge the battery sufficiently with the Osseo 100+ charger and try the automatic connection again. If this does not resolve the issue, please contact an authorized NSK dealer.
Higiena i konserwacja
iCare
- ACause1The power lead is not plugged in.Remedy1Check the connection.Cause2The power lead is not connected to the iCare.Remedy2Check the connection.Cause3The iCare is not switched ON.Remedy3Turn ON the black power switch (right side of iCare).Cause4The internal fuse has blown.Remedy4Replace the fuse.
- ACause1Oiling ports (1-4) are not switched on (the Connector LED is OFF).Remedy1Press one or more of the oiling ports.Cause2The door is OPEN before or during operation.Remedy2Close the door.Cause3The door is CLOSED before Chuck Lubrication.Remedy3Open the door before Chuck Lubrication.Cause4Oil chamber is empty (Error LED (Oil) lights up).Remedy4Fill with NSK iCare maintenance oil.Cause5Air is not supplied. Or the air pressure is low (Error LED (Air Pressure) lights up).Remedy5Check air supply and air pressure.Cause6The room temperature is too low.Remedy6The iCare will operate in temperatures of 10 - 40°C. Especially in winter, wait until the room temperature reaches the operating temperature.
- ACauseThe hole of the Chuck Lubricate Nozzle is clogged with debris.RemedyRemove dirt in the hole using a brush (Do not use a metal brush).
